Profile

Settings

Loading…

← Papers

Help

FAQ

Which features require signing in?

You can read the commentary, paper info, Model Structure and related papers of public papers without signing in. Sign in to manage your account or contact us about your inquiries.

Can I upload a PDF file?

Upload papers with an arXiv or DOI link; the commentary is public to everyone. Papers whose source and terms of use can't be verified, so the commentary can't be made public, can't be uploaded. Uploading PDF files and keeping papers private are for subscribers only, and a paper uploaded as a file is saved as private only. Searching for official info by title runs only when you press the button.

Are the figures and AI commentary accurate?

Figures come only from the original paper, and only those whose terms of use are verified are shown. The commentary is written by AI and may be wrong. Check important claims, numbers and figure context against the evidence in the original paper.

I want to report a copyright issue.

Terms of use and the source of the original paper are on the Paper info tab. If there's a problem, choose "Paper content, figures and copyright" under Contact us and send the paper link, the part in question, and evidence that you hold the rights. Receiving your report doesn't confirm any rights or remove anything right away. You can see the result of the review under My inquiries.

I forgot my password.

If you signed up with email, use Forgot password in the sign-in window to reset it with your verified email. If you signed up with Google, Naver or Kakao, change your password with that service.

What should I include in an inquiry?

Tell us which screen you were on and what you were doing when the problem happened, along with the paper link. Don't include personal information such as passwords, API keys, verification codes or phone numbers. Inquiries and replies are stored in Refivu and you can see them under My inquiries.

How do I delete my account?

In Settings, under My account, press Delete account. Review what will be deleted and complete identity verification to delete your account. For other requests, choose "Account" under Contact us.

Paper commentary

Design and Validation of a Robotic Knee Exoskeleton with Musculoskeletal Simulation-Derived Biological Torque Control

A knee exoskeleton trained entirely in simulation transfers to real hardware

  • Problem Biological torque controllers need data recorded on the target device, so every new device requires fresh data collection.
  • Method The authors train the estimator on knee moments from a muscle-driven squat simulation, so no device data is recorded, then run it on onboard sensors alone.
  • Results In seven participants, assistance cut peak biological knee extension moment by 21.5% during squatting compared with no exoskeleton.

Figure 5: Peak biological knee extension moment and muscle activation
with and without the exoskeleton.(a) Peak biological knee extension
moment, (b) peak vast…

Fig. 5 .Peak biological knee extension moment and muscle activation with and without the exoskeleton.(a) Peak biological knee extension moment, (b) peak vastus medialis activation, and (c) peak medial gastroc- nemius activation without (No Exo) and with (Exo) the exoskeleton. Bars represent group means, and error bars represent±1 standard deviation across participants. Asterisks indicate a significant difference between conditions (p <0.05, pairedt-test).

Original paper · CC BY 4.0 · Rajiv V. Joshi, Jimin An, Ilseung Park, Dawit Lee, Inseung Kang · Design and Validation of a Robotic Knee Exoskeleton with Musculoskeletal Simulation-Derived Biological Torque Control · Original paper figure · Extracted the figure area from page 5 of the original PDF.

Why device-bound estimators stall reuse

Gathering those datasets takes time and laboratory equipment, and it has to be repeated for every new device.

Open-source projects have released pieces of the exoskeleton stack, from orthoses and mechanical designs to firmware and trained controllers. Those pieces were not built to work together, so a controller still needs fresh data once the parts are assembled. Simulation and dataset reuse could shrink that burden, but no current exoskeleton approach pairs open hardware with a pipeline other groups can reproduce from simulation to controller.

Training labels from a muscle-driven squat simulation

The simulated imitation policy supplies the training labels, so the estimator never learns from data recorded on the device. The filtered joint trajectories are differentiated, which gives the joint velocities that the policy tracks.

Following the DeepMimic framework, the policy trained with soft actor-critic learns to reproduce the reference squat in this model. It outputs muscle excitations and a lumbar target angle, and a proportional-derivative controller tracks that angle. The policy runs at 100 Hz in episodes of up to 10 s, starting from sampled reference states.

An episode ends early if the pelvis deviates from the reference, so the policy cannot drift into unrealistic squats.

The reward combines three tracking terms for joint coordinates, joint velocities, and foot position with two penalty terms. The tracking weights are 0.60 for joint coordinates, 0.15 for joint velocities, and 0.25 for foot position. The penalties on mean squared muscle activation and step-to-step input changes push the policy toward plausible muscle coordination.

Simulated knee muscle moments from these squats become the regression targets for the estimator, scaled by a fixed normalization factor. The estimator sees only sensor-equivalent signals recorded during deterministic rollouts, because the policy itself relies on quantities the exoskeleton cannot measure.

Distilling the moment into a sensor-only estimator

The imitation policy depends on muscle states, contact forces, and future reference motion that the exoskeleton can't measure, so a separate estimator is trained on sensor-equivalent signals. The estimator learns from deterministic rollouts of the policy, sampled at each 100 Hz step. Each sample is therefore a three-channel vector, so the estimator sees the same signals the device carries.

A temporal convolutional network maps a 95-sample window, or 0.95 s, of these inputs to the current moment estimate.

Before training, the authors standardize the inputs. They add Gaussian noise to the gyroscope channels and offsets to the knee angle, so the estimator sees sensor noise and calibration error. The checkpoint with the lowest validation loss is kept as the final estimator.

On the exoskeleton, the embedded computer runs the estimator at 100 Hz on the onboard IMU and encoder signals. Only the gain k and the delay Δt are tuned for each participant during familiarization, and the output is clamped to 18 Nm. The delay also sets when peak torque occurs, so the data can't separate estimator timing from that tuning.

Keeping the sensor inputs clean in the knee exoskeleton hardware

The exoskeleton uses three cuffs per leg, at the thigh, shank, and ankle, to carry the knee's extension torque. The bilateral device weighs 3.5 kg, including the battery and embedded electronics. An earlier prototype put cuffs only on the lateral side, so extension torque passed through shear alone. At higher torques, the resulting transverse-plane moment rotated the device around the limb and corrupted the signals the controller reads.

The current design wraps each cuff around the anterior and posterior surfaces, so shear and normal force share the load.

The AK80-9 V3.0 quasi-direct-drive actuator sits at the knee, with an integrated motor driver and encoder. Its low gear ratio keeps the joint backdrivable, so the user moves with little impedance when no torque is commanded. The actuator's rated and peak torque are 9 and 22 Nm, but the stock motor driver limits output to 18 Nm. The waist-mounted Jetson Orin Nano runs the 100 Hz high-level controller and polls both microcontrollers each cycle.

The round trip between the tiers takes under 0.5 ms, leaving most of each cycle for the controller output.

After any communication fault, the joint returns to a passive, backdrivable state. A 22.2V lithium polymer battery worn at the waist powers both tiers, with an estimated run time of about 2 h. Because the structural parts are fabricated in-house, the authors could test revised cuffs with the same controller without retraining.

The estimator reads only the IMU and encoder signals, so cuff rotation feeds straight into its inputs. The estimator runs on 100 Hz inputs, so the hardware must deliver clean signals at that rate.

How the squatting trials were compared

Seven healthy male participants, aged 26.3±2.6 years, with body mass 76.5±8.7 kg and height 1.77±0.04 m, squatted in a motion capture laboratory. A set of 24 markers at anatomical landmarks fed inverse kinematics in OpenSim, with marker and ground reaction data sampled at 100 Hz. Surface EMG from the right leg covered the vastus medialis and the medial gastrocnemius, and each muscle had three MVC recordings.

Each participant first squatted for 60 s without the exoskeleton, paced at 40 beats per minute with a metronome.

Participants then did 5 to 8 familiarization squats, followed by a further 5-minute rest. The assisted trial used the same protocol as the baseline trial, so the two conditions were recorded under matched instructions. Trigger pulses from the embedded computer marked the start and end of each trial, aligning the exoskeleton data with motion capture.

That 3.5 kg was distributed as 0.90 kg on the pelvis, 0.91 kg on each thigh, and 0.39 kg on each shank, with inertia scaled to the added mass. Biological knee moment in assisted trials is the inverse-dynamics knee moment minus the aligned commanded torque. Each participant's final 10 complete cycles were time-normalized and averaged, excluding the initial adaptation period.

EMG was resampled from 2148 Hz to 1000 Hz before filtering and rectification. The MVC reference for each muscle was the maximum of a 250 ms moving average, and EMG was expressed as %MVC against that reference. Peak knee flexion, peak biological knee extension moment, and peak vastus medialis and medial gastrocnemius activation were extracted from each participant's mean squat-cycle profile.

Two-sided paired t-tests compared the conditions, and the authors set significance at p<0.05.

Knee moment and muscle effort under assistance

A paired t-test gave p=0.006, and the decrease appeared in all seven participants. The 17.6 Nm drop in biological moment sits close to that peak command, so the device seems to take over part of the effort.

The paired t-test gave p=0.016, and six of the seven participants showed the decrease. Because the biological moment was computed from commanded torque, the authors treat this EMG drop as independent support for the reduced effort. The authors read this flat ankle response as no detectable transfer of effort to the plantarflexors.

Because the knee angle followed a similar flexion pattern in both conditions, the comparison keeps the movement shape constant. The 7.4 degree shallower squat may account for part of the moment reduction. The group-averaged biological knee moment was lower in magnitude with the exoskeleton across the cycle, as Fig. 4 shows.

Limitations of the squatting-only evidence

The study tested only squatting, so the estimator's behavior in other tasks is still unknown. Walking and sit-to-stand need different inter-joint coordination, and the authors haven't shown the estimator carries over to them. Retraining the estimator on new reference motions is also untested.

The biological moment was computed from commanded torque, not from measured torque. Any transmission or interface losses would therefore inflate the estimated assistive contribution. The authors argue that these reductions likely understate the effect of assistance alone [15]. The fixed condition order may also have introduced fatigue or learning effects.

Table II compares devices from other studies, tasks and conditions, so it isn't a controlled comparison. Yu et al. reported a 38.1% vastus medialis reduction at 54 Nm during squatting. Boon et al. reported 26% at 55 Nm during stair ascent, a different task. The paper states that this study's 26.7% reduction matches Boon et al., though the tasks differ.

The authors propose the exoskeleton as a shared testbed for separating a controller's contribution from its hardware's. That testbed role depends on validation across these tasks and populations.

Paper info

Authors
Rajiv V. Joshi, Jimin An, Ilseung Park, Dawit Lee, Inseung Kang
Venue
arXiv · Preprint
First published
2026-10-30
Added
2026-11-7
Updated
2026-10-30
Version
v1
arXiv
2610.41981
Category
cs.RO
Original paper terms
CC BY 4.0
What we show
Paper info, commentary and permitted figures from the original paper

AI commentary is a separate explanation from the original paper and may contain errors. The original paper and figures follow each source's terms of use and any rights noted separately.

Abstract

Data-driven controllers have enabled lower-limb robotic exoskeletons to adapt assistance to the user across many tasks. A leading example is biological torque control, which estimates the user's own joint moment from onboard sensors and assists in proportion to that estimate. However, its estimators are trained on data recorded with the target device, so deploying the controller on new hardware requires collecting new data. Existing open-source efforts release hardware, software, or controllers separately, so this data collection must still be repeated on the assembled device. Here, we present an open-source knee exoskeleton, built from off-the-shelf components and widely available fabrication tools, with a biological torque controller derived entirely from simulation. A muscle-driven imitation policy reproduces squatting motions from an open biomechanics dataset, and the resulting knee moments are distilled into an estimator that uses only onboard sensors, so no training data are recorded on the device. In seven able-bodied participants, assistance reduced peak biological knee extension moment by 21.5% and peak vastus medialis activation by 26.7% during squatting, relative to squatting without the exoskeleton (p<0.05). Our results demonstrate that a biological torque controller derived from simulation can transfer to physical hardware without device-specific data collection. Because the hardware and training pipeline are released together, research groups can train and compare knee exoskeleton controllers on shared hardware without additional data collection overhead.

Figures from the paper

Quote from the paper

Rajiv V. Joshi, Jimin An, Ilseung Park, Dawit Lee, Inseung Kang (2026). Design and Validation of a Robotic Knee Exoskeleton with Musculoskeletal Simulation-Derived Biological Torque Control. arXiv:2610.41981v1. https://arxiv.org/abs/2610.41981v1

BibTeX
@misc{joshi2026design,
  title = {{Design and Validation of a Robotic Knee Exoskeleton with Musculoskeletal Simulation-Derived Biological Torque Control}},
  author = {Rajiv V. Joshi and Jimin An and Ilseung Park and Dawit Lee and Inseung Kang},
  year = {2026},
  eprint = {2610.41981},
  archivePrefix = {arXiv},
  url = {https://arxiv.org/abs/2610.41981v1}
}